Detecto 7045G Spécifications Page 51

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 153
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 50
5.2.3. Setup
5.2.4. Test Procedure
The test procedure was in accordance with EN 61000-4-3
a) The testing was performed in a fully anechoic chamber. The transmit antenna was located at a
distance of 3 meters from the EUT.
b) The frequency range is swept from 80 MHz to 1000 MHz, with the signal 80% amplitude modulated
with a 1kHz sine-wave. The rate of sweep did not exceed 1.5 x 10 -3 decade/s, where the frequency
range is swept incrementally, the step size was 1% of preceding frequency value.
c) The dwell time at each frequency shall be not less than the time necessary for the EUT to be able to
respond.
d) The test was performed with the EUT exposed to both vertically and horizontally polarized fields on
each of the four sides.
A Test Lab Techno Corp.
Report Number: 1402CE13-01
51 of 153
Vue de la page 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 152 153

Commentaires sur ces manuels

Pas de commentaire